GPIB Bus Compatible Interface - The general purpose interface bus (GPIB) is a stan-
dard bus for parallel control of electronic equipment. The two meter (6.6 foot) cable
supplied connects the Calorimeter unit to a user-supplied GPIB controller. The GPIB
cable meets the requirements of IEEE Standard 488-1978.
Accessory Items A two meter (6.6 ft) GPIB cable is supplied with the Calorimeter. Other GPIB cables
are available as follows:
Recommended Test
Equipment
Table 1 list equipment recommended for performance testing and maintenance of the
Calorimeter. The critical equipment parameters listed are the minimum specifica-
tions applicable to test and measurement devices other than those listed. Valid Calo-
rimeter tests and measurements are critical and depend largely on the accuracy of
the test and measurement devices used. Refer to Chapter 5 - Maintenance for more
information on testing and calibration.
Electromagnetic
Interference
The Calorimeter has been tested and was found to meet Class A limits as specified in
Part 15J of the FCC requirement for electromagnetic emissions.
3
Introduction
Bird Part Number Cable Length
5-1317-1 1 meter (3.3 ft)
5-1317-3 4 meter (13.1 ft)
Equipment Type Recommended Model Critical Equipment Parameters
AC Power Source Elgar 501SL with PIP 9012 Output - 10 to 200 watts @ 50
ohms
Stability - greater than 0.1%
THD - less than 0.25%
Frequency - adjustable to 60, ±1 Hz
AC Power Meter YEW 2533 Range - 10 to 200 watts
Accuracy - 0.1%
Automatic Network
Analyzer
HP 8753 with 85046A ±0.2 dB accuracy for return loss
measurements
Cable Assembly Test connections as shown in figure
12 in Chapter 5 - Maintenance
Table 1
Recommended
Test Equipment