Bird-technologies 5015 Series-Manual Manuale Utente Pagina 23

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 31
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 22
11
Chapter 4 Specifications
5011 Specifications
Power Measurement Characteristics
General Terminated average power
Frequency Range 40 MHz to 4 GHz
Power Measurement
Range
–20.0 to +10.0 dBm (10 µW to 10
mW)
Maximum Power 2 W avg., 125 W peak for 5 µs
Peak/Average Ratio 12 dB maximum
Accuracy ± (5% of reading
a
+ 1.0 µW)
(excluding mismatch uncertainty)
No correction factors necessary
a. Above 40 °C, when measuring frequencies between 40
and 100 MHz, add 1%.
Input Impedance 50 Ohms (nominal)
Input VSWR:
Typical
Maximum
1.03 (36.6 dB return loss)
1.20 (20.8 dB return loss)
Input Connector Precision N Male
Output Connector Male DB-9 to host instrument
Power Supply From host instrument via cable
connection
Vedere la pagina 22